diff --git a/p6_gas-network-total-failure-backup-plan_bt1-2/p6_gas-network-total-failure-backup-plan_bt1-2.md b/p6_gas-network-total-failure-backup-plan_bt1-2/p6_gas-network-total-failure-backup-plan_bt1-2.md new file mode 100644 index 0000000000000000000000000000000000000000..588861e2735f12d8b659ff710d3f1475d992ef1c --- /dev/null +++ b/p6_gas-network-total-failure-backup-plan_bt1-2/p6_gas-network-total-failure-backup-plan_bt1-2.md @@ -0,0 +1,49 @@ +--- +layout: page +permalink: /internal/lab/p6_gas-network-total-failure-backup-plan_bt1-2/ +--- + +# P6: Gas network total failure backup plan BT1/BT2 + +## CO2 +This process covers the actions to take in case of total failure of the carbon dioxide (CO2, dioxyde de carbone) gas network due to pipeline infrastructure breakdown. + +1. Check the CO2 connection at the back of each incubator and at the connection of the manometer to identify a possible leak. +To do so, use the leak detector spray or the gas leak detector device. + +2. Also investigate the LSRU floors + +3. If a leak can be identified, fix it: + +- plug the tubing back +- close the manometer connecting the incubator to the network (see arrow on the following picture): transfer the cells to another incubator + +<div align="center"> +<img src="p6_img_1.png"> +</div> + +4. If there no leak is detected, call 5555 and ask them to contact the SIL duty call. + +5. Notify the LCSB duty call manager and their backup by email, fill and fill the [Duty call report form](https://ulsurvey.uni.lu/index.php/936567?lang=en) + +## N2 + +This process covers the actions to take in case of total failure of the nitrogen (N2, azote) gas network due to pipeline infrastructure breakdown. +It consists in the rescue of glove boxes and hypoxic incubators by connecting them to bottles of nitrogen. + +1. Identify the pipeline infrastructure breakdown: + +- Check the “gas bottle exchange record sheet†(Appendix 03) located on the door inside the gas deposit room. +- If the N2 bottles have been changed in the last 48 h, this is indicative of a pipeline infrastructure break down. + +2. Check the N2 connection at the back of each glovebox and hypoxic incubator and at the connection of the manometer to identify a possible leak. +To do so, use the leak detector spray or the gas leak detector device. + +3. Also investigate the LSRU floors +4. If a leak can be identified, fix it: + +- plug the tubing back +- close the manometer, do not transfer the samples + +5. If there no leak is detected, call 5555 and ask them to contact the SIL duty call. +6. Notify the LCSB duty call manager and their backup by email, fill and fill the [Duty call report form](https://ulsurvey.uni.lu/index.php/936567?lang=en) \ No newline at end of file diff --git a/p6_gas-network-total-failure-backup-plan_bt1-2/p6_img_1.png b/p6_gas-network-total-failure-backup-plan_bt1-2/p6_img_1.png new file mode 100644 index 0000000000000000000000000000000000000000..19cae341a69afeda8991d6e4e076bd9d9ca5b5ec Binary files /dev/null and b/p6_gas-network-total-failure-backup-plan_bt1-2/p6_img_1.png differ