 # P3: Incubator failure
@@ -13,8 +17,8 @@ This process applies in case of problem with temperature or gas (N2 or CO2) leve
 3. If it is not the problem, come on site to evaluate the situation.
 4. If the issue cannot be fixed on site, transfer the samples to another incubator in the same room with the same temperarure and gas setting.
 
-Personal protective equipment  
--Lab coat  
+Personal protective equipment
+-Lab coat
 -Gloves
 
 5. Keep track of what was moved where.
diff --git a/p.8.2_cryostorage_bta/p.8.2_cryostorage_bta.md b/p.8.2_cryostorage_bta/p.8.2_cryostorage_bta.md
index 46f938ec0d1e7088ce884e4e8c7be14509ee6a21..edddaa23ed3676779aa0c0adfb2532b03295f29f 100644
--- a/p.8.2_cryostorage_bta/p.8.2_cryostorage_bta.md
+++ b/p.8.2_cryostorage_bta/p.8.2_cryostorage_bta.md
@@ -1,15 +1,19 @@
 ---
 layout: page
 permalink: /internal/lab/p.8.2_cryostorage_bta/
+shortcut: lab:p.8.2_cryostorage_bta
+redirect_from:
+  - /cards/lab:p.8.2_cryostorage_bta
+  - /internal/cards/lab:p.8.2_cryostorage_bta
 ---
 
-# P8.2: Cryostorage BTA  
-1.	If you receive an alarm for low levels of liquid nitrogen, do not come on site.  
+# P8.2: Cryostorage BTA
+1.	If you receive an alarm for low levels of liquid nitrogen, do not come on site.
 
 2.	The only case where you need to come is in the event of a temperature rise above - 80°C (however, if the nitrogen tank is empty, the temperature should be kept for a week.
-There is no need to move the samples)  
+There is no need to move the samples)
 
-3.	If this happens, transfer the samples to the -80°C freezer  
+3.	If this happens, transfer the samples to the -80°C freezer
 
 4.	Notify the Facility Manager, the LCSB duty call manager and their backup by email and fill the [Duty call report form](https://ulsurvey.uni.lu/index.php/936567?lang=en)
 
diff --git a/p6_gas-network-total-failure-backup-plan_bt1-2/p6_gas-network-total-failure-backup-plan_bt1-2.md b/p6_gas-network-total-failure-backup-plan_bt1-2/p6_gas-network-total-failure-backup-plan_bt1-2.md
index 588861e2735f12d8b659ff710d3f1475d992ef1c..906fc79e864d1d0fcbf6127b25f5104e747e30a3 100644
--- a/p6_gas-network-total-failure-backup-plan_bt1-2/p6_gas-network-total-failure-backup-plan_bt1-2.md
+++ b/p6_gas-network-total-failure-backup-plan_bt1-2/p6_gas-network-total-failure-backup-plan_bt1-2.md
@@ -1,49 +1,53 @@
 ---
 layout: page
 permalink: /internal/lab/p6_gas-network-total-failure-backup-plan_bt1-2/
+shortcut: lab:p6_gas-network-total-failure-backup-plan_bt1-2
+redirect_from:
+  - /cards/lab:p6_gas-network-total-failure-backup-plan_bt1-2
+  - /internal/cards/lab:p6_gas-network-total-failure-backup-plan_bt1-2
 ---
 
 # P6: Gas network total failure backup plan BT1/BT2
 
-## CO2  
-This process covers the actions to take in case of total failure of the carbon dioxide (CO2, dioxyde de carbone) gas network due to pipeline infrastructure breakdown.  
+## CO2
+This process covers the actions to take in case of total failure of the carbon dioxide (CO2, dioxyde de carbone) gas network due to pipeline infrastructure breakdown.
 
 1. Check the CO2 connection at the back of each incubator and at the connection of the manometer to identify a possible leak.
-To do so, use the leak detector spray or the gas leak detector device.  
+To do so, use the leak detector spray or the gas leak detector device.
 
-2.	Also investigate the LSRU floors  
+2.	Also investigate the LSRU floors
 
 3.	If a leak can be identified, fix it:
 
 - plug the tubing back
-- close the manometer connecting the incubator to the network (see arrow on the following picture): transfer the cells to another incubator  
+- close the manometer connecting the incubator to the network (see arrow on the following picture): transfer the cells to another incubator
 
 <div align="center">
 <img src="p6_img_1.png">
-</div>  
+</div>
 
-4. If there no leak is detected, call 5555 and ask them to contact the SIL duty call.  
- 
-5. Notify the LCSB duty call manager and their backup by email, fill and fill the [Duty call report form](https://ulsurvey.uni.lu/index.php/936567?lang=en)  
+4. If there no leak is detected, call 5555 and ask them to contact the SIL duty call.
 
-## N2  
+5. Notify the LCSB duty call manager and their backup by email, fill and fill the [Duty call report form](https://ulsurvey.uni.lu/index.php/936567?lang=en)
+
+## N2
 
 This process covers the actions to take in case of total failure of the nitrogen (N2, azote) gas network  due to pipeline infrastructure breakdown.
-It consists in the rescue of glove boxes and hypoxic incubators by connecting them to bottles of nitrogen.  
+It consists in the rescue of glove boxes and hypoxic incubators by connecting them to bottles of nitrogen.
 
-1. Identify the pipeline infrastructure breakdown:  
+1. Identify the pipeline infrastructure breakdown:
 
-- Check the “gas bottle exchange record sheet” (Appendix 03) located on the door inside the gas deposit room.  
-- If the N2 bottles have been changed in the last 48 h, this is indicative of a pipeline infrastructure break down.  
+- Check the “gas bottle exchange record sheet” (Appendix 03) located on the door inside the gas deposit room.
+- If the N2 bottles have been changed in the last 48 h, this is indicative of a pipeline infrastructure break down.
 
 2. Check the N2 connection at the back of each glovebox and hypoxic incubator and at the connection of the manometer to identify a possible leak.
-To do so, use the leak detector spray or the gas leak detector device.  
+To do so, use the leak detector spray or the gas leak detector device.
 
-3. Also investigate the LSRU floors  
-4. If a leak can be identified, fix it:  
+3. Also investigate the LSRU floors
+4. If a leak can be identified, fix it:
 
-- plug the tubing back  
-- close the manometer, do not transfer the samples  
+- plug the tubing back
+- close the manometer, do not transfer the samples
 
-5. If there no leak is detected, call 5555 and ask them to contact the SIL duty call.  
-6. Notify the LCSB duty call manager and their backup by email, fill and fill the [Duty call report form](https://ulsurvey.uni.lu/index.php/936567?lang=en) 
\ No newline at end of file
+5. If there no leak is detected, call 5555 and ask them to contact the SIL duty call.
+6. Notify the LCSB duty call manager and their backup by email, fill and fill the [Duty call report form](https://ulsurvey.uni.lu/index.php/936567?lang=en)
\ No newline at end of file
diff --git a/p9.2_intervention-for-emergency-situations_bta/p9.2_intervention-for-emergency-situations_bta.md b/p9.2_intervention-for-emergency-situations_bta/p9.2_intervention-for-emergency-situations_bta.md
index e2e5702122acfb90a4c2a8402cec78d0def5482a..565e335d5d798f1130a4e58c63a084eab106ff9b 100644
--- a/p9.2_intervention-for-emergency-situations_bta/p9.2_intervention-for-emergency-situations_bta.md
+++ b/p9.2_intervention-for-emergency-situations_bta/p9.2_intervention-for-emergency-situations_bta.md
@@ -1,14 +1,18 @@
 ---
 layout: page
 permalink: /internal/lab/p9.2_intervention-for-emergency-situations_bta/
+shortcut: lab:p9.2_intervention-for-emergency-situations_bta
+redirect_from:
+  - /cards/lab:p9.2_intervention-for-emergency-situations_bta
+  - /internal/cards/lab:p9.2_intervention-for-emergency-situations_bta
 ---
 
-# P9.2: Intervention for emergency situations BTA  
+# P9.2: Intervention for emergency situations BTA
 
-This procedure has to be applied in the event of any problem with the animal facility, major water leak, fire or ventilation issue.  
+This procedure has to be applied in the event of any problem with the animal facility, major water leak, fire or ventilation issue.
 
-1.	Call the PC security (+352 46 66 44 5555) and explain the situation and ask them to contact the SIL duty call.  
+1.	Call the PC security (+352 46 66 44 5555) and explain the situation and ask them to contact the SIL duty call.
 
-2. If necessary, request the intervention of public rescue (112 for firemen and ambulances; 113 for the police) and be available when they arrive on site.  
+2. If necessary, request the intervention of public rescue (112 for firemen and ambulances; 113 for the police) and be available when they arrive on site.
 
 3.	Notify the Facility Manager, the LCSB duty call manager and their backup by email and fill the [Duty call report form](https://ulsurvey.uni.lu/index.php/936567?lang=en).
\ No newline at end of file