After the data extraction is finished, the quality of the retrieved
spectra can be checked using the criteria defined in the
_Prescreening_ tab of the _Configuration_ Section. The resulting
_summary table_ is given below and is based on the prescreening
criteria and filter and ordering setup specified in _Filter and order
the summary table_ configuration subsection.
After the automatic quality control check based on
the criteria defined in the _Prescreening_ tab of the _Configuration_ Section is done , the resulting
_summary table_ is given below to manually update the quality of the retrieved spectra.
To update the sumary table,
* Click any row for MS1 table and the corresponding MS2 peaks are shown on the right pane.
* Manual QA can be updated by double clicking on the required QA fields namely _MS1/MS2 match?_, _Above threshold_, _MS2 exists_ after inspecting the peaklist.
* To save the changes, click the _Commit manual QA_ tab and the summary table can be exported as _CSV_ file by clicking the _Export MS1/MS2 summary_ tab.
* For any given ID, the MS1 and MS2 fragments can be visualized by selecting the corrsponding row in the table.
* The user can zoom-in by selection the reion of interest using the cursor and double click to zoom-out.
*The user can select multiple rows to display different plots and allows the user to save individual plot by clicking _save this plot_ tab or save all the plots.