# Utilization of balances
## Routine maintenance of analytical balances from Mettler-Toledo XS4002S and XPE205DR
### 1. Personal protective equipment
* Always wear a lab coat and gloves when using the balances.
* If necessary, wear safety glasses.
* If weighing dangerous, volatile compounds, use a balance located under a chemical hood.
There are such balances on the 5th floor in BT1 and on the 2nd floor in BT2.
### 2. Weighing
**XPE205DR: used for 1.4 mg to 220 g**
<div align="center">
<img src="XPE205DR.jpg" height="400">
</div>
**XS4002S: used for 800 mg to 4.1 kg**
<div align="center">
<img src="XS4002S.jpg" height="400">
</div>
1. For XPE balances, check that the air bubble is in the middle of the circle.
It is located on the rear right foot of the instrument.
<div align="center">
<img src="XPE205DR_2.jpg" height="400">
</div>
2. Switch on the balance by pressing on “On/Off” until the display appears.
3. In order to allow an equilibration of electronics within the balance, allow it to stand for 10 to 15 minutes before starting using it.
4. Zero the balance by pressing on “0”.
a. For the XPE balance, make sure the doors are closed by pressing on “↨” or by passing your hand over the right and left upper parts of the terminal.
5. Place an anti-static weighing boat or a weighing paper in the middle of the pan of the balance. Do not place the material to be weighted directly on the pan.
6. If you weigh directly in a tube, place the Ergo Clip on the pan of the balance and place your tube in it.
**Ergo clip**:
<div align="center">
<img src="Ergo-Clip_2.jpg" height="400">
</div>
7. Tare the balance by pressing on “T”
8. Place the material to be weighted in the anti-static weighing boat or on the weighing paper in the middle of the pan to avoid corner-load error, for the XPE balance,close the doors after loading your material.
9. To prevent contamination of stock material, do not return unused substance to the stock bottle.
10. After use, switch the balance off by pressing on the “ON/OFF” button.
11. Contact the Quality Assistant if the logbook is full.
12. If you notice anything wrong with the instrument, contact the instrument responsible.
### 3. Cleaning
1. After each use, the balance and the material used to weight have to be cleaned.
2. **XPE205DR**:
* Open the doors and remove the weighing pan
* Wipe the surfaces with a brush
* Wipe the weighing pan and the inside of the chamber with a damp cloth
* Place the weighing pan back and make sure it is correctly placed
3. **XS4002S**:
* Wipe the surfaces with a brush
* Wipe the weighing pan with a damp cloth
* If some powder fell off the pan, carefully remove the crown around the pan and clean it to remove any residue. If you need assistance, contact the instrument responsible person
<img src="Crown.jpg" height="400">
* When the crown is removed, carefully wipe with a damp cloth around the weighing pan
* Carefully place the crown back
* Always wipe the area around the balance with 70 % ethanol to remove any substance
* Spoons: wash them with a mild detergent and water
\ No newline at end of file
utilization-of-pH-meter/description_pH-meters.jpg

51.4 KiB

---
layout: page
permalink: /internal/utilization-of-pH-meter/
---
# Utilization of pH meter
## Routine Maintenance of pH meter “Seven Easy” and “Seven Compact” from Mettler-Toledo.
## 1. Description
<div align="center">
<img src="img/../description_pH-meters.jpg" height="300">
</div>
## 2. Personal protective equipment
Always wear a lab coat and gloves when working with the pH meter. Wear safety glasses if necessary.
## 3. Prepare the calibration reagents
1. The calibration cannot be performed using the reagents directly in their original containers, they need to be aliquoted.
2. Transfer a few millilitres of each of the three calibration solutions in a 15 mL falcon tube. Label them with the pH value of the solution and the date of aliquot.
3. Indicate the opening date on the original bottles and store them in the dark.
4. Keep the aliquoted solutions for a month, then replace them with fresh ones.
5. Indicate the date of exchange of the solutions in the log book.
## 4. Calibration
Calibration should be performed only if it has not been performed on the same day (see logbook) or if you have doubts about the calibration status of the instrument.
The pH meter allows for calibration of 1, 2 or 3 points.
The calibration standard solutions are of pH 4, 7 and 10.
1. Rinse the electrode with deionized water, wipe carefully the outside of the electrode with a lint free tissue.
2. Place the electrode in a calibration standard and press “CAL”.
3. The pH meter automatically endpoints when calibrating.
4. For 2 points calibration, repeat steps 1 to 3 for the second calibration standard.
5. For 3 points calibrations, repeat steps 1 to 3 with the third calibration standard.
6. Rinse the electrode with deionized water, wipe carefully the outside of the electrode with a lint free tissue.
7. After calibration, to return to sample measurement, press on:
a. “Read” for the Seven Easy
b. “End”, then “Save” for the Seven Compact
8. Add the calibration date in the equipment logbook.
## 5. Measuring the pH
1. pH is temperature sensitive, therefore, make sure that your solution is at the temperature you will use it (for example: 37°C for cell culture medium).
2. Make sure the sample has been thoroughly mixed.
a. If you use magnetic stirrer, make sure you stop the stirrer before inserting the probe.
3. Rinse the electrode with deionized water, wipe carefully the outside of the electrode with a lint free tissue.
4. Place the electrode in the sample and press on “READ” to start the measurement.
5. During measurement, the decimal point of the display flashes.
6. Once the measurement is complete, the display freezes automatically.
7. Rinse the electrode with deionized water and place it back in its storing solution container (a 15 mL falcon containing 5 mL of 3 M KCl for example).
8. Adjust the pH of your solution, use a Pasteur glass pipet if you use strong acids or bases.
## 6. Maintenance
1. All calibration solutions should be kept away from the light.
2. Any spillage has to be cleaned immediately. Use wipes to dry the spill, then clean with water.
3. The electrode has to be maintained in 3 M KCl.
a. Do not allow it to dry.
b. Change the solution at least monthly.
c. Indicate the date of change in the log book.
d. If the filling solution is encrusted outside the electrode, remove it by rinsing the electrode with deionized water.
4. Contact the quality assistant if the logbook is full.
5. Would you notice any issue with the instrument, create a ticket.