@@ -12,14 +12,13 @@ This section describes the procedure in the event of low nitrogen levels in both
1. If the issue is not the nitrogen level and the temperature cannot be held below -120°C, transfer the content of the cryostorage to the back up -80°C freezer LCSB0709 in the same room, see “P6: Freezer failure back up plan”.
<divalign="center">
|*Personal protective equipment*|
|:-----------------------------:|
|Lab coat|
|Cryogloves|
|Safety goggles|
</div>
2. If the issue comes from the nitrogen level, verify whether the cryocyl-120 tanks is empty by pressing on “press” on the level gauge panel: in that case, 0% is displayed