P3: Incubator failure
This process does not concern LCSB0814 and LCSB1047 in BT2, 3rd floor, room 310 (See "P4: Zebrafish incubator temperature problem"). This process applies in case of problem with temperature or gas (N2 or CO2) levels of an incubator.
- If after 20 minutes, the trend does not go back to the initial settings.
- If you suspect an open door, you can ask a security guard to close it (+352 46 66 44 5555).
- If it is not the problem, come on site to evaluate the situation.
- If the issue cannot be fixed on site, transfer the samples to another incubator in the same room with the same temperarure and gas setting.
Personal protective equipment
-Lab coat
-Gloves
- Keep track of what was moved where.
- Notify the LCSB duty call manager and their back up by email, and fill the "Duty call report form" (https://ulsurvey.uni.lu/index.php/936567?lang=eng)