......@@ -16,9 +16,13 @@ This maintenance is important to preserve the integrity of the glass bottles ins
To ensure proper functioning of the centrivap concentrators and associated cold trap, a routine maintenance is necessary:
<div align=center>
**_Personal Protective Equipment_**:
* Lab coat
* Nitrile gloves
Lab coat
Nitrile gloves
</div>
1. This routine maintenance is performed by the users of the instrument.
......@@ -39,10 +43,14 @@ To drain, use the faucet on the left side of the cold trap.
**Once a year, the refrigeration system condenser of the Cold trap should be cleaned.**
<div align=center>
**_Personal Protective Equipment_**:
* Lab coat
* Nitrile gloves
* Safety goggles
Lab coat
Nitrile gloves
Safety goggles
</div>
1. This step is performed by a maintenance officer
......@@ -77,4 +85,4 @@ When necessary, the acid cartridge (LIMS ID: 10608) should be exchanged
* Place the old cartridge in the chemical hood for evaporation in a retention container
* 24 hours later, trash it in a blue bin UN3291
\ No newline at end of file
* 24 hours later, trash it in a blue bin UN3291