# P8.2: Cryostorage BTA
1. If you receive an alarm for low levels of liquid nitrogen, do not come on site.
2. The only case where you need to come is in the event of a temperature rise above - 80°C (however, if the nitrogen tank is empty, the temperature should be kept for a week.
There is no need to move the samples)
3. If this happens, transfer the samples to the -80°C freezer
4. Notify the Facility Manager, the LCSB duty call manager and their backup by email and fill the [Duty call report form](https://ulsurvey.uni.lu/index.php/936567?lang=en)