Utilization of balances
Routine maintenance of analytical balances from Mettler-Toledo XS4002S and XPE205DR
Personal protective equipment
- Always wear a lab coat and gloves when using the balances.
- If necessary, wear safety glasses.
- If weighing dangerous, volatile compounds, use a balance located under a chemical hood.
There are such balances on the 5th floor in BT1 and on the 2nd floor in BT2.
Weighing
XPE205DR: used for 1.4 mg to 220 g
XS4002S: used for 800 mg to 4.1 kg
- For XPE balances, check that the air bubble is in the middle of the circle. It is located on the rear right foot of the instrument.
- Switch on the balance by pressing on “On/Off” until the display appears.
- In order to allow an equilibration of electronics within the balance, allow it to stand for 10 to 15 minutes before starting using it.
- Zero the balance by pressing on “0”. a. For the XPE balance, make sure the doors are closed by pressing on “↨” or by passing your hand over the right and left upper parts of the terminal.
- Place an anti-static weighing boat or a weighing paper in the middle of the pan of the balance. Do not place the material to be weighted directly on the pan.
- If you weigh directly in a tube, place the Ergo Clip on the pan of the balance and place your tube in it.
Ergo clip:
- Tare the balance by pressing on “T”
- Place the material to be weighted in the anti-static weighing boat or on the weighing paper in the middle of the pan to avoid corner-load error, for the XPE balance,close the doors after loading your material.
- To prevent contamination of stock material, do not return unused substance to the stock bottle.
- After use, switch the balance off by pressing on the “ON/OFF” button.
- Contact the Quality Assistant if the logbook is full.
- If you notice anything wrong with the instrument, contact the instrument responsible.
Cleaning
- After each use, the balance and the material used to weight have to be cleaned.
-
XPE205DR:
- Open the doors and remove the weighing pan
- Wipe the surfaces with a brush
- Wipe the weighing pan and the inside of the chamber with a damp cloth
- Place the weighing pan back and make sure it is correctly placed
-
XS4002S:
- Wipe the surfaces with a brush
- Wipe the weighing pan with a damp cloth
- If some powder fell off the pan, carefully remove the crown around the pan and clean it to remove any residue. If you need assistance, contact the instrument responsible person
* When the crown is removed, carefully wipe with a damp cloth around the weighing pan * Carefully place the crown back * Always wipe the area around the balance with 70 % ethanol to remove any substance * Spoons: wash them with a mild detergent and water