# P3: Incubator failure # P3: Incubator failure
...@@ -13,8 +17,8 @@ This process applies in case of problem with temperature or gas (N2 or CO2) leve ...@@ -13,8 +17,8 @@ This process applies in case of problem with temperature or gas (N2 or CO2) leve
3. If it is not the problem, come on site to evaluate the situation. 3. If it is not the problem, come on site to evaluate the situation.
4. If the issue cannot be fixed on site, transfer the samples to another incubator in the same room with the same temperarure and gas setting. 4. If the issue cannot be fixed on site, transfer the samples to another incubator in the same room with the same temperarure and gas setting.
Personal protective equipment Personal protective equipment
-Lab coat -Lab coat
-Gloves -Gloves
5. Keep track of what was moved where. 5. Keep track of what was moved where.
......
--- ---
layout: page layout: page
permalink: /internal/lab/p.8.2_cryostorage_bta/ permalink: /internal/lab/p.8.2_cryostorage_bta/
shortcut: lab:p.8.2_cryostorage_bta
redirect_from:
- /cards/lab:p.8.2_cryostorage_bta
- /internal/cards/lab:p.8.2_cryostorage_bta
--- ---
# P8.2: Cryostorage BTA # P8.2: Cryostorage BTA
1. If you receive an alarm for low levels of liquid nitrogen, do not come on site. 1. If you receive an alarm for low levels of liquid nitrogen, do not come on site.
2. The only case where you need to come is in the event of a temperature rise above - 80°C (however, if the nitrogen tank is empty, the temperature should be kept for a week. 2. The only case where you need to come is in the event of a temperature rise above - 80°C (however, if the nitrogen tank is empty, the temperature should be kept for a week.
There is no need to move the samples) There is no need to move the samples)
3. If this happens, transfer the samples to the -80°C freezer 3. If this happens, transfer the samples to the -80°C freezer
4. Notify the Facility Manager, the LCSB duty call manager and their backup by email and fill the [Duty call report form](https://ulsurvey.uni.lu/index.php/936567?lang=en) 4. Notify the Facility Manager, the LCSB duty call manager and their backup by email and fill the [Duty call report form](https://ulsurvey.uni.lu/index.php/936567?lang=en)
--- ---
layout: page layout: page
permalink: /internal/lab/p6_gas-network-total-failure-backup-plan_bt1-2/ permalink: /internal/lab/p6_gas-network-total-failure-backup-plan_bt1-2/
shortcut: lab:p6_gas-network-total-failure-backup-plan_bt1-2
redirect_from:
- /cards/lab:p6_gas-network-total-failure-backup-plan_bt1-2
- /internal/cards/lab:p6_gas-network-total-failure-backup-plan_bt1-2
--- ---
# P6: Gas network total failure backup plan BT1/BT2 # P6: Gas network total failure backup plan BT1/BT2
## CO2 ## CO2
This process covers the actions to take in case of total failure of the carbon dioxide (CO2, dioxyde de carbone) gas network due to pipeline infrastructure breakdown. This process covers the actions to take in case of total failure of the carbon dioxide (CO2, dioxyde de carbone) gas network due to pipeline infrastructure breakdown.
1. Check the CO2 connection at the back of each incubator and at the connection of the manometer to identify a possible leak. 1. Check the CO2 connection at the back of each incubator and at the connection of the manometer to identify a possible leak.
To do so, use the leak detector spray or the gas leak detector device. To do so, use the leak detector spray or the gas leak detector device.
2. Also investigate the LSRU floors 2. Also investigate the LSRU floors
3. If a leak can be identified, fix it: 3. If a leak can be identified, fix it:
- plug the tubing back - plug the tubing back
- close the manometer connecting the incubator to the network (see arrow on the following picture): transfer the cells to another incubator - close the manometer connecting the incubator to the network (see arrow on the following picture): transfer the cells to another incubator
<div align="center"> <div align="center">
<img src="p6_img_1.png"> <img src="p6_img_1.png">
</div> </div>
4. If there no leak is detected, call 5555 and ask them to contact the SIL duty call. 4. If there no leak is detected, call 5555 and ask them to contact the SIL duty call.
5. Notify the LCSB duty call manager and their backup by email, fill and fill the [Duty call report form](https://ulsurvey.uni.lu/index.php/936567?lang=en)
## N2 5. Notify the LCSB duty call manager and their backup by email, fill and fill the [Duty call report form](https://ulsurvey.uni.lu/index.php/936567?lang=en)
## N2
This process covers the actions to take in case of total failure of the nitrogen (N2, azote) gas network due to pipeline infrastructure breakdown. This process covers the actions to take in case of total failure of the nitrogen (N2, azote) gas network due to pipeline infrastructure breakdown.
It consists in the rescue of glove boxes and hypoxic incubators by connecting them to bottles of nitrogen. It consists in the rescue of glove boxes and hypoxic incubators by connecting them to bottles of nitrogen.
1. Identify the pipeline infrastructure breakdown: 1. Identify the pipeline infrastructure breakdown:
- Check the “gas bottle exchange record sheet” (Appendix 03) located on the door inside the gas deposit room. - Check the “gas bottle exchange record sheet” (Appendix 03) located on the door inside the gas deposit room.
- If the N2 bottles have been changed in the last 48 h, this is indicative of a pipeline infrastructure break down. - If the N2 bottles have been changed in the last 48 h, this is indicative of a pipeline infrastructure break down.
2. Check the N2 connection at the back of each glovebox and hypoxic incubator and at the connection of the manometer to identify a possible leak. 2. Check the N2 connection at the back of each glovebox and hypoxic incubator and at the connection of the manometer to identify a possible leak.
To do so, use the leak detector spray or the gas leak detector device. To do so, use the leak detector spray or the gas leak detector device.
3. Also investigate the LSRU floors 3. Also investigate the LSRU floors
4. If a leak can be identified, fix it: 4. If a leak can be identified, fix it:
- plug the tubing back - plug the tubing back
- close the manometer, do not transfer the samples - close the manometer, do not transfer the samples
5. If there no leak is detected, call 5555 and ask them to contact the SIL duty call. 5. If there no leak is detected, call 5555 and ask them to contact the SIL duty call.
6. Notify the LCSB duty call manager and their backup by email, fill and fill the [Duty call report form](https://ulsurvey.uni.lu/index.php/936567?lang=en) 6. Notify the LCSB duty call manager and their backup by email, fill and fill the [Duty call report form](https://ulsurvey.uni.lu/index.php/936567?lang=en)
\ No newline at end of file \ No newline at end of file
--- ---
layout: page layout: page
permalink: /internal/lab/p9.2_intervention-for-emergency-situations_bta/ permalink: /internal/lab/p9.2_intervention-for-emergency-situations_bta/
shortcut: lab:p9.2_intervention-for-emergency-situations_bta
redirect_from:
- /cards/lab:p9.2_intervention-for-emergency-situations_bta
- /internal/cards/lab:p9.2_intervention-for-emergency-situations_bta
--- ---
# P9.2: Intervention for emergency situations BTA # P9.2: Intervention for emergency situations BTA
This procedure has to be applied in the event of any problem with the animal facility, major water leak, fire or ventilation issue. This procedure has to be applied in the event of any problem with the animal facility, major water leak, fire or ventilation issue.
1. Call the PC security (+352 46 66 44 5555) and explain the situation and ask them to contact the SIL duty call. 1. Call the PC security (+352 46 66 44 5555) and explain the situation and ask them to contact the SIL duty call.
2. If necessary, request the intervention of public rescue (112 for firemen and ambulances; 113 for the police) and be available when they arrive on site. 2. If necessary, request the intervention of public rescue (112 for firemen and ambulances; 113 for the police) and be available when they arrive on site.
3. Notify the Facility Manager, the LCSB duty call manager and their backup by email and fill the [Duty call report form](https://ulsurvey.uni.lu/index.php/936567?lang=en). 3. Notify the Facility Manager, the LCSB duty call manager and their backup by email and fill the [Duty call report form](https://ulsurvey.uni.lu/index.php/936567?lang=en).
\ No newline at end of file